ANALYSIS OF EXPOSURE INDEX AND DEVIATION INDEX IN AEC MODE FOR RADIOGRAPHIC IMAGE QUALITY IMPROVEMENT AT THUNG SONG HOSPITAL
Main Article Content
Abstract
This study aimed to 1) Analyze and compare the Exposure Index (EI) and Deviation Index (DI) values obtained from two digital radiography machines using the Automatic Exposure Control (AEC) mode. 2) Investigate the relationship between EI and DI with imaging parameters (kVp and mAs), and 3) Establish optimal parameter settings for the AEC system to ensure consistent image quality and patient safety. This experimental research utilized a total of 62 chest posteroanterior (PA) radiographs acquired in AEC mode, retrieved from the hospital’s PACS. The dataset included 31 images from each of the two machines and additional measurements from a Phantom chest model. EI and DI values were extracted from PACS, and image quality was assessed by three radiologists using a 5-point Likert scale. Data were analyzed using independent t-tests and regression analysis. The results revealed that 1) There was no statistically significant difference in EI values between the two machines (p = 0.109), while DI values showed a significant difference (p < 0.00001), with Machine B producing DI values closest to zero. 2) A low correlation was observed between EI and kVp (r = 0.227), and a very weak negative correlation between DI and mAs (r = -0.114). 3) Machine B, operating at 110 kVp and 3.56 mAs, yielded DI values nearest to zero and the highest image quality scores, indicating a more stable AEC performance. These findings highlight the necessity of establishing machine-specific standard values for EI and DI to ensure consistent image quality, reduce unnecessary radiation exposure, and enhance diagnostic accuracy.
Article Details
References
Prasad, K. N., Cole, W. C., & Haase, G. M. (2004). Radiation protection in humans: extending the concept of as low as reasonably achievable (ALARA) from dose to biological damage. The British journal of radiology, 77(914), 97-99.
Harrison, J. D., Balonov, M., Bochud, F., Martin, C., Menzel, H. G., Ortiz-Lopez, P., Wakeford, R. (2021). ICRP publication 147: use of dose quantities in radiological protection. Annals of the ICRP, 50(1), 9-82.
Doyle, P., & Martin, C. J. (2006). Calibrating automatic exposure control devices for digital radiography. Physics in Medicine & Biology, 51(21), 5475.
Seibert, J. A., & Morin, R. L. (2011). The standardized exposure index for digital radiography: an opportunity for optimization of radiation dose to the pediatric population. Pediatric radiology, 41, 573-581.
Mothiram, U., Brennan, P. C., Robinson, J., Lewis, S. J., & Moran, B. (2013). Retrospective evaluation of exposure index (EI) values from plain radiographs reveals important considerations for quality improvement. Journal of medical radiation sciences, 60(4), 115-122.
Luechabhun, C., Prasertdumrongchai, S., Pukleeb, B., Suwanpradit, P., & Apipunyasopon, L. (2024). Optimization of exposure techniques for lumbosacral spine radiographic examination: a phantom study. The Thai Journal of Radiological Technology, 49(1), 103-109.
Chanlumpoo, N. (2019). การตั้งค่าเทคนิคการให้ปริมาณรังสีที่เหมาะสมสำหรับการสร้าง ภาพรังสีระบบดิจิทัลในหุ่นจำลอง. Thammasat University Hospital Journal Online, 4(1), 14-30.
Lewis, S., Pieterse, T., & Lawrence, H. (2019). Evaluating the use of exposure indicators in digital x-ray imaging system: Gauteng South Africa. Radiography, 25(3), e58-e62.
Moonkum, N., Jitchom, S., Sukaram, S., Nimtrakool, N., Boonrat, P., & Tochaikul, G. (2023). Determination of scattered radiation dose for radiological staff during portable chest examinations of COVID-19 patients. Radiological Physics and Technology, 16(1), 85-93.
ประภัสสร ไกรหาญ. (2021). การประเมินค่าดัชนีชี้วัดปริมาณรังสี (Exposure index: EI, Deviation index: DI) ในการถ่ายภาพทรวงอก เด็ก อายุ 0-12 ปี โรงพยาบาลสิริน ธร จังหวัดขอนแก่น. วารสารสิ่งแวดล้อมศึกษาการแพทย์และสุขภาพ, 6(4), 18-27.
แสงชัย สมนาค, กฤษณา ร้อยศรี, และรังสี ทรงประโคน. (2024). ค่าปริมาณรังสีอ้างอิง สำหรับการถ่ายภาพรังสีวินิจฉัยทางการแพทย์ในโรงพยาบาลสุรินทร์. วารสารการแพทย์โรงพยาบาลศรีสะเกษ สุรินทร์ บุรีรัมย์, 39(1), 163-173.
Huda, W., & Abrahams, R. B. (2015). X-ray-based medical imaging and resolution. American Journal of Roentgenology, 204(4), W393-W397.
International Commission on Radiological Protection. (2019). Radiological protection in digital radiology (ICRP Publication 135). ICRP.
Seibert, J. A., Bogucki, T., Ciona, T., Huda, W., Karellas, A., Mercier, J., Samei, E., Shepard, J., Stewart, B., Strauss, K., Suleiman, O., Tucker, D., Uzenoff, R., Weiser, J., & Willis, C. (2006). Acceptance Testing and Quality Control of Photostimulable Storage Phosphor Imaging Systems. AAPM.
Seetasung, S. (2024). ผลของการปรับค่าเทคนิคการถ่ายภาพรังสีต่อค่าดัชนีเบี่ยงเบนของ ปริมาณรังสีและค่าปริมาณรังสีอ้างอิงจากภาพถ่ายรังสีทรวงอกแบบดิจิทัลในโรงพยาบาลสมเด็จพระยุพราช หล่มเก่า. The Thai Journal of Radiological Technology, 49(1), 110-120